High-Quality 50.4W Medical Power Supplies with Interchangeable Plugs from China Suppliers and Factory
`
`
Energy storage system
Product Specification
| OUTPUT | Voltage(DC) | 9V | 9.1-14V | 14.1-21V | 21.1-30V | 30.1-48V |
| Current(Max.) | 5A | 4.16A | 3.36A | 2.3A | 1.57A | |
| Power (Max.) | 45W | 50W | 50.4W | 50.4W | 50.4W | |
| Ripple&noise | ≤150 mV | ≤200 mV | ≤200 mV | ≤300 mV | ≤300 mv | |
| voltage tolerance | ±5% | |||||
| Standby power | <0.21W | |||||
| INPUT | Voltage Range | 100-240VAC | ||||
| Frequency Range | 50-60Hz | |||||
| ENVIROMENT | operation Temp. | 0°C to +40°C | ||||
| Storage Temp. | -20°C to +80°C | |||||
| Operation HUM | 10%-90% | |||||
| SAFETY | Safety Standards | 60601 CB, cULus, FCC, CE, TUV Mark, AU plug report |
||||
| Other | MTBF | >50,000 hours at 80% load and 25°C | ||||
| Dimension | 84.5X52.0X35.0mm | |||||
Energy storage system
Specification Details
1. Inrush Current: Cold start inrush current will not exceed 80A at 100-240Vac input and Max load at 25°C.
2. Average Efficiency: More than 87.40% at 115Vac and 230Vac input (tested at 25%, 50%, 75%, and 100% of max load).
3. No Load Power Loss: Less than 0.1W when input is 115Vac or 230Vac and output is no load.
4. Capacitance Load: Adapter turns on normally with 100~240Vac input and 470uF capacitance load, output in rated range.
5. Transient Response: Output voltage will not exceed 10% of nominal voltage (load fluctuation: 20% to 80%, R/S: 0.5A/uS, frequency: 100Hz, duration: 8mS at 80%).
6. Over Current Protection: 3.6A < I < 8A. Output enters hiccup mode under over-current, self-recovers after fault removal.
7. Short Circuit Protection: Input power decreases during output short circuit. Power supply remains undamaged and self-recovers.
8. Over Voltage Protection: Power supply triggers protection mode when output exceeds rated voltage.
9. Vibration Test: 10 to 300Hz sweep at constant acceleration of 1.0G (Breadth: 3.5mm) for 1 hour per perpendicular X, Y, Z axis.
10. Drop Test: Height: 1m; fell 3 times on a plank of at least 53mm thickness placed on non-flexible cement ground.
11. EMI Compliance: FCC Part18 class B, EN55032.
12. EMS Compliance: EN55035.
13. Safety Accordance: ANSI/AAMI ES60601-1, CSA C22.2 NO. 60601-1, IEC60601-1, EN60601-1, AS/NZS 60601.1.
ELITE
Why Choose US?
Quality Assurance Process
- We perform rigorous Turn ON/Off tests at rated load and 100V/240V input.
- 100% of products undergo full functional testing: DC output performance, current ripple, efficiency, inrush current, RMS current, peak current, input power, harmonic violation, power factor, short circuit, low voltage, over current, over voltage, over load, and over temperature protection.
- 100% of products undergo a 4-hour rated load burn-in test at normal input and 80% rated load under 40°C±5°C.
- Comprehensive final appearance inspection is performed on all units.
Frequently Asked Questions
What is the input voltage and frequency range for this system?
The system supports a wide input voltage range of 100-240VAC with an input frequency range of 50-60Hz, making it compatible with global power standards.
What safety certifications does this energy storage system hold?
It complies with major international safety standards, including ANSI/AAMI ES60601-1, CSA C22.2 NO. 60601-1, IEC60601-1, EN60601-1, and AS/NZS 60601.1. It also holds CB, cULus, FCC, CE, and TUV Mark certifications.
How is product quality guaranteed before shipment?
We ensure quality through 100% functional testing (including output performance, protections, and efficiency), a 4-hour rated load burn-in test at 40°C±5°C, and a thorough final appearance inspection.
What protective safety features are built into the device?
The power supply features comprehensive safety protections, including Over Current Protection (hiccup mode with self-recovery), Short Circuit Protection (self-recovery without damage), Over Voltage Protection, Low Voltage Protection, Over Load Protection, and Over Temperature Protection.
What are the operating temperature and environment limits?
The system operates reliably in temperatures from 0°C to +40°C and relative humidity levels of 10%-90%. It can be safely stored in temperatures ranging from -20°C to +80°C.
What is the expected lifespan (MTBF) of the system?
The Mean Time Between Failures (MTBF) is rated at greater than 50,000 hours when operating under an 80% load at an ambient temperature of 25°C.