| OUTPUT | Voltage(DC) | 9V | 9.1-14V | 14.1-21V | 21.1-30V | 30.1-48V |
| Current(Max.) | 5A | 4.16A | 3.36A | 2.3A | 1.57A | |
| Power (Max.) | 45W | 50W | 50.4W | 50.4W | 50.4W | |
| Ripple & Noise | ≤150 mV | ≤200 mV | ≤200 mV | ≤300 mV | ≤300 mv | |
| Voltage Tolerance | ±5% | |||||
| Standby Power | <0.21W | |||||
| INPUT | Voltage Range | 100-240VAC | ||||
| Frequency Range | 50-60Hz | |||||
| ENVIRONMENT | Operation Temp. | 0℃ to +40℃ | ||||
| Storage Temp. | -20℃ to +80℃ | |||||
| Operation HUM | 10%-90% | |||||
| SAFETY | Safety Standards | 60601, CB, cULus, FCC, CE, TUV Mark, AU plug report | ||||
| OTHER | MTBF | >50,000 hours at 80% load and 25℃ | ||||
| Dimension | 84.5 X 52.0 X 35.0 mm | |||||
Technical Features
Inrush Current: Cold start inrush current will not exceed 80A at 100-240Vac input and Max load at 25℃.
Efficiency: Average efficiency is more than 87.40% at 115Vac and 230Vac input (tested at 25%, 50%, 75%, and 100% load).
No-Load Power: Input power loss is less than 0.1W when output is at no load (115Vac/230Vac).
Capacitance Load: Adapter turns on normally with a capacitance load of 470uF at 100~240Vac.
Transient Response: Output fluctuation within 10% of nominal voltage (20% to 80% load change, 0.5A/uS).
Over Current Protection: Hiccup mode at 3.6A<I<8A with self-recovery.
Short Circuit & Over Voltage: Full protection with automatic recovery and no damage during faults.
Durability: Passed vibration tests (1.0G, 10-300Hz) and 1m drop tests on solid ground.
Compliance: EMI (FCC Part18 class B, EN55032), EMS (EN55035), and global safety standards (IEC60601-1, etc.).
Why Choose US?
Quality Assurance Protocols
Strict Testing Procedures:
- Functional Tests: DC output, Current ripple, Efficiency, Inrush current, RMS current, Peak current, Power factor, and Harmonics.
- Protection Verification: Short circuit, Low voltage, Over current, Over voltage, Over load, and Over temperature protection tests.
- Burn-in Test: 100% of products undergo a 4-hour rated load burn-in test at 40℃±5℃.
- Final Inspection: Comprehensive appearance and On/Off testing at 100V/240V inputs.
Frequently Asked Questions
1. What is the input voltage range for this energy storage system?
The system supports a wide universal input range of 100-240VAC at a frequency of 50-60Hz, making it compatible with global power standards.
2. Does the power supply include safety certifications?
Yes, it complies with multiple safety standards including ANSI/AAMI ES60601-1, IEC60601-1, and has certifications such as CB, cULus, FCC, CE, and TUV Mark.
3. How is the product's reliability tested?
Every single unit undergoes a 100% functional test and a 4-hour burn-in test at 40℃±5℃ under 80%-100% rated load to ensure long-term stability.
4. What kind of protection features are built-in?
The system includes Over Current Protection (hiccup mode), Short Circuit Protection, Over Voltage Protection, and Over Temperature Protection with self-recovery capabilities.
5. What is the standby power consumption?
The power supply is designed for high efficiency with a standby power consumption of less than 0.21W.
6. Can the system handle physical stress like vibrations or drops?
Yes, it has passed vibration tests (10 to 300Hz at 1.0G) and a 1-meter drop test on solid surfaces to ensure durability during transport and use.